Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 5720A: Correction File Initialization DATE: 23-Jan-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 1 NUMBER OF LINES: 28 ============================================================================= # # Compatibility: # MET/CAL 4.23 or later # # Required Files: # CORRECT.EXE Correction file utility program. # 5720.NEW 5720A correction file without any correction data. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP This procedure copies the file "5720.new" to "5720.cor" 1.001 DISP in the MET/CAL correction file directory wiping out all 1.001 DISP previous correction data including DC Amps and AC Volts. 1.001 DISP If 5720.cor already exists it is backed up as 5720.bak. 1.001 DISP 1.001 DISP You must run this procedure at least once before running 1.001 DISP the 3458A verification for the first time. 1.002 DOS correct 5720 init 1.003 END